Digital Medical X-ray System Specification |
|
No. |
Parameters and Functions: |
1.1 |
Clinical application: Perform X-ray imaging examination of standing position, lying position and horizontal side position of various parts of the human body, realize X-ray digital imaging, digital image DICOM network transmission, printing, storage management and laser-printable film, perfect image Post processing function; |
2 |
Product Technical Specifications and Requirements |
2.1 |
X-Ray Tube |
2.1.1 |
Anode heat capacity: 220KHU |
2.1.2 |
Focus: small focus 0.6mm, large focus 1.2mm |
2.1.3 |
Focus power: small focus 20KW, large focus 50KW |
2.2 |
High Voltage Generator |
2.2.1 |
Output power: 50KW |
2.2.2 |
Input line power: 380V, 50Hz |
2.2.3 |
Output voltage range: 40~150KV |
2.2.4 |
Minimum exposure time: 1ms; |
2.2.5 |
Output current range: 10~630mA |
2.3 |
Flat Panel Detector |
2.3.1 |
Effective acquisition area: 43×43cm |
2.3.2 |
Data transmission: wired |
2.3.3 |
Pixel size: 139 μm |
2.3.4 |
Pixel matrix: 3072×3072 |
2.3.5 |
Limit resolution: 3.7 lp/mm |
2.3.6 |
The system can meet the dual-panel models. |
2.4 |
Tube Assembly Support |
2.4.1 |
The tube assembly support structure: column type |
2.4.2 |
Column, gantry and diagnostic table are integrated; |
2.4.3 |
The tube support arm can be extended or rotated |
2.4.4 |
The tube supported arm can be motorized along the column; Range: 90cm; |
2.4.5 |
The tube is rotated by ±90° along the support arm; The column can be rotated ±180° along the vertical axis. |
2.5 |
Vertical Chest Stand |
2.5.1 |
The vertical move by motor diver, range of the detection center: 135cm; |
2.5.2 |
Configurable detachable movable grid; |
2.6 |
Diagnostic table |
2.6.1 |
Tabletop can float in 4 directions and can be electromagnetically locked |