Digital Medical X-ray System Specification | |
No. | Parameters and Functions: |
1.1 | Clinical application: Perform X-ray imaging examination of standing position, lying position and horizontal side position of various parts of the human body, realize X-ray digital imaging, digital image DICOM network transmission, printing, storage management and laser-printable film, perfect image Post processing function; |
2 | Product Technical Specifications and Requirements |
2.1 | X-Ray Tube |
2.1.1 | Anode heat capacity: 220KHU |
2.1.2 | Focus: small focus 0.6mm, large focus 1.2mm |
2.1.3 | Focus power: small focus 20KW, large focus 50KW |
2.2 | High Voltage Generator |
2.2.1 | Output power: 50KW |
2.2.2 | Input line power: 380V, 50Hz |
2.2.3 | Output voltage range: 40~150KV |
2.2.4 | Minimum exposure time: 1ms; |
2.2.5 | Output current range: 10~630mA |
2.3 | Flat Panel Detector |
2.3.1 | Effective acquisition area: 43×43cm |
2.3.2 | Data transmission: wired |
2.3.3 | Pixel size: 139 μm |
2.3.4 | Pixel matrix: 3072×3072 |
2.3.5 | Limit resolution: 3.7 lp/mm |
2.3.6 | The system can meet the dual-panel models. |
2.4 | Tube Assembly Support |
2.4.1 | The tube assembly support structure: column type |
2.4.2 | Column, gantry and diagnostic table are integrated; |
2.4.3 | The tube support arm can be extended or rotated |
2.4.4 | The tube supported arm can be motorized along the column; Range: 90cm; |
2.4.5 | The tube is rotated by ±90° along the support arm; The column can be rotated ±180° along the vertical axis. |
2.5 | Vertical Chest Stand |
2.5.1 | The vertical move by motor diver, range of the detection center: 135cm; |
2.5.2 | Configurable detachable movable grid; |
2.6 | Diagnostic table |
2.6.1 | Tabletop can float in 4 directions and can be electromagnetically locked |
2.6.2 | Tabletop longitudinal movement range: 900mm |
2.6.3 | Tabletop lateral movement range: 250mm |
2.7 | Integrated Imaging Acquisition and Processing Workstation |
2.7.1 | CPU: 3.3GHz; OS: Windows 7; RAM: 8GB; Hard disk: 1T; |
2.7.2 | DICOM 3.0 output standard; integrated image acquisition, image processing, image output and case management functions; |